Nova™ Microcinerator Operating Instructions Consignes d' utilisation Instrucciones de operaciones Bedienungsanleitung © 2005 Argos Technolog
10Trouble Shooting Switch not illuminated – Check Fuse and replace if required. Burner not heating – If the burner is not heating and glowing re
11Replacement Parts Ordering Information Replacement parts should be ordered from your local distributor Nova™ Microcinerator Part No. Descripti
12 Accessories 3 Place Slide Dryer – Part Number S3012. Use the heat already generated by your sterilizer to dry your slides. Aluminium Lo
13 The Slide Dryer and Loop Rest can be used together.
14Repair Service Should your Nova™ Microcinerator require service or repair, please contact Argos Technologies customer service. They will issue a
15Replacing the Sterilizing Element Important – Only qualified personnel should carry out repairs to this product. Always switch the unit “off” an
16Tools required: • Cross headed (Philips) screwdriver • 7mm AF (9/32” AF) spanner or socket wrench 1. Remove screws (1). 2. Remove front face
2 Contents Page Subject 2 Product Description and Environmental Conditions 3 Materials and Specifications 4 Symbols Used 5 Warranty Informati
3 Materials and Specification Voltages Materials: Part No S3010 - 100V, 50 HZ, 235W, 2.3 amps Stainless Steel 316 Part No S3011 – 120V, 60 HZ
4Symbols Used T A Easily touched high temperature parts Time Lag Fuse Amperes Protective Conductor Terminal IMPORTANT: If this produ
5 WARRANTY INFORMATION The Nova™ Microcinerator is guaranteed free from defects in workmanship and materials for a period of one year after the cu
6 Part Descriptions © Argos Technologies Inc 2005 Note: to ensure optimum performance your Nova™ Microcinerator should be left on all day. Durin
7Sterilizing Loops and Needles 1. Carefully insert the loop or needle at least halfway into the sterilizing element to avoid aerosoling any residu
8Tilting the Burner Head The burner head angle can be altered to suit your working style. Allow the burner to cool for 1 hour before adjusting. T
9Maintenance, Care & Safety Precautions WARNING!! The Nova™ Microcinerator should never be submerged in water or any other cleaning compounds.